Introduction
CO2-M01 is used to match Patient Monitor. The module adopts sidestream method to measure the parameters related to ETCO2 and AWRR. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow and indirectly reflect the alveolar ventilation.
The application population is adult patients. It is applicable for use in emergency department, ICU, operating room, respiratory department and other occasions. It is connected to the nose end of the patient through the sampling tube (applied part).
Features
1)End-tidal CO2 measurement function.
2)Inspired Minimum CO2 measurement function.
3)Respiration rate measurement function.
4)Optional three units: %, kPa, mmH.
5)Be compatible with Respironics module's parameters and protocol.
6)No uesr calibration is required.
7)User can calibrate zero by themselves.
8)With atmospheric pressure and temperature compensation functions.
Performance
End-tidal CO2 measurement:
1)Measurement range: 0mmHg ~150mmHg;
2) Resolution: 1 mmHg;
3)Accuracy:
0~40 mmHg: ±2 mmHg
41~70 mmHg: ±5% of readings
71~100 mmHg: ±8% of readings
101~150 mmHg: ±10% of readings
Respiration rate measurement:
1)Measurement range: 2rpm~120rpm;
2)Resolution: 1rpm;
3)Accuracy: ±1rpm;
Inspired Minimum CO2 measurement:
Measurement range: 3mmHg ~50mmHg;
Other parameters:
1)Preheat time: 2min;
2)Response time: < 3s;
3)Sampling frequency: 200 Hz.
Safety characteristic
1)Degree of protection against electric shock: type BF applied part;
2)Degree of protection against ingress of liquid: IP22.
Accessory
1)A uers manual;
2)A CO2 sampling tube(nasal cannula with filter cotton).
Physical Characteristic
Dimension: 98.9*54.3*27.6 mm
Weight: 165 g
Working environment
Temperature: 5ºC~40ºC
Relative humidity: 30%~ 75%, no condensation
Atmospheric pressure: 700hPa~1060hPa
Power supply: DC 5.0V
Input power: ≤0.9VA