Introduction
The module adopts mainstream method to measure the parameters related to ETCO2 and AWRR. It can detect pulmonary ventilation function and reflect the circulation and pulmonary blood flow condition,and indirectly reflect the alveolar ventilation, which is applicable for use in emergency department, ICU, operating room, respiratory department and other occasions for adult and pediatric patients. The module connects with patient' respiratory tract by the air adapter (applied part) and breathing tube. The breathing tube except for air adapter is not a part of the device.
Features
1)End-tidal CO2 measurement function;
2)Inspired Minimum CO2 measurement function;
3)Respiration rate measurement function;
4)Optional three units: %, kPa, mmHg;
5)Be compatible with Respironics module's parameters and protocol;
6)No user calibration is required;
7)Ueser can calibrate zero by themselves;
8)Use with the adult / pediatric air adapter to achieve mainstream sampling;
9)With atmospheric pressure and temperature compensation functions;
Performance
End-tidal CO2 measurement:
1)Measurement range: 0 ~150mmHg
2)Resolution: 1mmHg
3)Accuracy: 0~40 mmHg:±2 mmHg
41~70 mmHg:±5% of readings
71~100 mmHg:±8% of readings
101~150 mmHg:±10% of readings
Respiration rate measurement:
1)Measurement range: 2rpm~150rpm
2)Resolution: 1rpm
3)Accuracy: ±1rpm
Inspired Minimum CO2 measurement:
Measurement range: 3mmHg ~50mmHg;
Safety characteristic:
1)The degree of protection against electric shock: type BF applied part
Other parameters:
1)Preheat time: 2min;
2)Response time: < 3s;
3)Sampling frequency: 200 Hz.
Accessory
1)A uers manual;
2)Adult/Pediatric air adapter.
Physical Characteristic
Dimension: 59*19.9*31.9mm
Weight: 84g
Power supply: DC 5.0V
Input power: <0.9VA
Working current: ≤150mA
Working environment
Temperature: 0ºC~40ºC
Relative humidity: 10%~ 75%, no condensation
Atmospheric pressure: 400mmHg~800mmHg